Sudan Red (SUD) ELISA kit

Citations(0)Uniprot : N/A
  • Cat.No.: AE81052SF

  • Reactivity: N/A

To Purchase AE81052SF

Size:
48T 96T
Price: Inquire

Reagent Preparation







Results demonstration







Assay Procedure Summary







Certificate



Product Details

Species Reactivity N/A
UniProt N/A
Abbreviation SUD
Alternative Names N/A
Range 0.1-8.1 ppb
Sensitivity 0.1 ppb
Sample Type Tomato juice, tomato paste, chilli paste, chilli powder, egg, feed, etc
Detection Method Competitive ELISA
Analysis Method Quantitive
Assay Duration 1-3h
Sample Volume 10-200 μL
Detection Wavelengt 450 nm

Test principle


 

Product Overview


 

Components


Reagents

Quantity

Reagents

Quantity

Assay plate (96 Wells)

1

Instruction manual

1

Standards

6 x 1 mL

Redissolving Solution (concentrate 2 x)

2 x 20 mL

Antibody

1 x 6 mL

HRP-Conjugate

1 x 6 mL

Wash Buffer (concentrate 20 x)

1 x 20 mL

Stop Solution

1 x 6 mL

Substrate A

1 x 6 mL

Substrate B

1 x 6 mL

Adhesive Films

4

 

Specificity


 

Recovery


Sample Type

Number

Recovery range (%)

Average(%)

Serum

10

90-101

96

EDTA plasma

10

89-97

93

Heparin plasma

10

91-99

95

 

Precision

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. CV (%) = SD/meanX100 Intra-Assay: CV<8% Inter-Assay: CV<12%
 

Linearity


Sample Type

1:2

1:4

1:8

1:16

Serum

78-89%

81-99%

92-103%

95-105%

EDTA plasma

91-101%

90-98%

93-101%

91-98%

Heparin plasma

92-103%

93-102%

92-99%

91-101%

 

Stability

The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. The loss rate was determined by accelerated thermal degradation test. Keep the kit at 37°C for 4 and 7 days, and compare O.D.values of the kit kept at 37°C with that of at recommended temperature. (referring from China Biological Products Standard, which was calculated by the Arrhenius equation. For ELISA kit, 4 days storage at 37°C can be considered as 6 months at 2 - 8°C, which means 7 days at 37°C equaling 12 months at 2 - 8°C).
 

Sample collection and storage

Store at 2-8°C. Please refer to Instruction Manual.
 

Kits storage instructions

Store at 2-8°C. Please refer to Instruction Manual.