Diethylstilbestrol (DES) ELISA Kit

Citations(0)Uniprot : N/A
  • Cat.No.: AE81023SF

  • Reactivity: N/A

To Purchase AE81023SF

Size:
48T 96T
Price: Inquire

Reagent Preparation







Results demonstration







Assay Procedure Summary







Certificate



Product Details

Species Reactivity N/A
UniProt N/A
Abbreviation DES
Alternative Names N/A
Range 0.025-2.025 ppb
Sensitivity 0.025 ppb
Sample Type Urine, tissue, feed, etc.
Detection Method Competitive ELISA
Analysis Method Quantitive
Assay Duration 1-3h
Sample Volume 10-200 μL
Detection Wavelengt 450 nm

Test principle

This assay is based on the competitive enzyme immunoassay for the detection of DES in the sample. The conjugate antigen is pre-coated on the micro-well stripes. The DES in the sample competes with the conjugate antigen pre-coated on the micro-well stripes, to interact with the antibodies against DES. After the addition of the enzyme conjugate, the TMB substrate is added for coloration. The optical density (OD) value of the sample has a negative correlation with the content of DES in the sample. This value is compared to the standard curve and the DES concentration is subsequently obtained.
 

Product Overview

Diethylstilbestrol (DES, former BAN stilboestrol) is a synthetic nonsteroidal estrogen that was first synthesized in 1938. It is also classified as an endocrine disruptor. Human exposure to DES occurred through diverse sources, such as dietary ingestion from supplemented cattle feed and medical treatment for certain conditions, including breast and prostate cancers. From about 1940 to 1971, DES was given to pregnant women in the mistaken belief it would reduce the risk of pregnancy complications and losses. In 1971, DES was shown to cause a rare vaginal tumor in girls and women who had been exposed to this drug in utero. The United States Food and Drug Administration subsequently withdrew DES from use in pregnant women. Follow-up studies have indicated that DES also has the potential to cause a variety of significant adverse medical complications during the lifetimes of those exposed. The United States National Cancer Institute recommends women born to mothers who took DES undergo special medical exams on a regular basis to screen for complications as a result of the drug. Individuals who were exposed to DES during their mothers' pregnancies are commonly referred to as “DES daughters” and “DES sons”.
 

Components


Reagents

Quantity

Reagents

Quantity

Assay plate (96 Wells)

1

Instruction manual

1

Standard

6 x 1 mL

Redissolving Solution (concentrate 5 x)  

2 x 20 mL

Antibody

1 x 7 mL

HRP-Conjugate

1 x 12 mL

Wash Buffer (concentrate 20 x) 

2 x 20 mL

Stop Solution

1 x 7 mL

Substrate A

1 x 7 mL

Substrate B

1 x 7 mL

Adhesive Films

4




 

Specificity


 

Recovery


Sample Type

Number

Recovery range (%)

Average(%)

Serum

10

90-101

96

EDTA plasma

10

89-97

93

Heparin plasma

10

91-99

95

 

Precision

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. CV (%) = SD/meanX100 Intra-Assay: CV<8% Inter-Assay: CV<12%
 

Linearity


Sample Type

1:2

1:4

1:8

1:16

Serum

78-89%

81-99%

92-103%

95-105%

EDTA plasma

91-101%

90-98%

93-101%

91-98%

Heparin plasma

92-103%

93-102%

92-99%

91-101%

 

Stability

The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. The loss rate was determined by accelerated thermal degradation test. Keep the kit at 37°C for 4 and 7 days, and compare O.D.values of the kit kept at 37°C with that of at recommended temperature. (referring from China Biological Products Standard, which was calculated by the Arrhenius equation. For ELISA kit, 4 days storage at 37°C can be considered as 6 months at 2 - 8°C, which means 7 days at 37°C equaling 12 months at 2 - 8°C).
 

Sample collection and storage

Store at 2-8°C. Please refer to Instruction Manual.
 

Kits storage instructions

Store at 2-8°C. Please refer to Instruction Manual.